China High Quality Dc Tester Manufacturers – 30kVA.200kV Automatic AC.DC Hipot Test Set GDYD-2030A – HV Hipot

China High Quality Dc Tester Manufacturers – 30kVA.200kV Automatic AC.DC Hipot Test Set GDYD-2030A – HV Hipot

Brief Description:


Product Detail

Product Tags

Related Video

Feedback (2)

We persistently execute our spirit of ''Innovation bringing growth, Highly-quality making sure subsistence, Administration marketing reward, Credit history attracting clients for Surge Voltage Test, Capacitance Tan Delta Tester, Power Transformer Test Set, Our merchandise are widely recognized and trusted by users and can meet up with continuously changing economic and social needs.
China High Quality Dc Tester Manufacturers – 30kVA.200kV Automatic AC.DC Hipot Test Set GDYD-2030A – HV Hipot Detail:

Features

● 7inch color LCD screen, with PLC automation control.
● HV Voltage, LV voltage, LC current measured at the same time, with high precision sensor and high-performance collection chip.
● Automatic mode and manual mode for option.
● Display HV voltage, LV voltage, LV current, time, and withstanding result.
● Over-voltage and over-current protection. Output voltage, HV current upper limit, LV current upper limit and timing can be set.
● Zero-starting function. Test can be proceeded after the voltage is back to zero.
● Store up to 8000pcs data in 3months. It has USB port to upload data to USB disk.
● When setting voltage is arrived, it will be timing automatically. After timing, voltage will be back to zero.
● When the LV current is over setting value, the voltage will be cut off automatically and it will be back to zero. When the HV current is over setting value, the test object will be cut off and test can be proceeded continuously.
● With anti-interference ability, suitable for electromagnetic environment.
● With wheels for both control unit and HV unit, easy to move.

Technical Specifications

Control unit specifications
● Power: AC380V, 50/60Hz
● Capacity: 30kVA
● Output voltage: AC 0-430V
● Rated output current: 75A
● Rated current rectified range: 0-75A
● HV voltmeter range: 0-200kV
● Accuracy: 1.5% of the reading±1digit (F.S)
● HV ammeter range: 0-100mA
● Accuracy: 1.5% of the reading±1digit (F.S)
● LV ammeter range: 0-25A
● Accuracy: 2.0% of the reading±1digit (F.S)
● Timing range: 1s-99h
● Weight: about 116kg
● Dimension: 520mm (L)*460mm (W)*1140mm (H)

Testing transformer (HV unit) specifications
● Input voltage: 0-400V
● Input current: AC 0-75A
● Output voltage: AC 0-200kV, DC 0-200kV
● Output current: AC 150mA, DC 106mA
● Capacity: 30kVA
● HV side/LV side turns ratio: 500:1
● Measuring voltage: AC 0-100V
● Accuracy: 1.5% of the reading±1 digit (F.S)
● No-load current: <10%
● Impedance voltage: <10%
● Weight: about 180kg
● Dimension: 630mm(D)*1730mm(H)

Working conditions
● Working place: No gas, steam, chemical dust and explosive and corrosive medium which will influence insulation.
● Environment temperature: -20–50℃.
● Environment humidity: ≤85% RH (no condensing).


Product detail pictures:

China High Quality Dc Tester Manufacturers –  30kVA.200kV Automatic AC.DC Hipot Test Set GDYD-2030A – HV Hipot detail pictures

China High Quality Dc Tester Manufacturers –  30kVA.200kV Automatic AC.DC Hipot Test Set GDYD-2030A – HV Hipot detail pictures


Related Product Guide:

We believe that long time period partnership is a result of top of the range, value added services, rich expertise and personal contact for China High Quality Dc Tester Manufacturers – 30kVA.200kV Automatic AC.DC Hipot Test Set GDYD-2030A – HV Hipot , The product will supply to all over the world, such as: Los Angeles, Bulgaria, Boston, All the imported machines effectively control and guarantee the machining precision for the items. Besides, we have a group of high-quality management personnels and professionals, who make the high-quality items and have the ability to develop new merchandise to expand our market home and abroad. We sincerely expect customers come for a blooming business for both of us.
  • We have been cooperated with this company for many years, the company always ensure timely delivery ,good quality and correct number, we are good partners.
    5 Stars By Mandy from Frankfurt - 2018.12.28 15:18
    The supplier cooperation attitude is very good, encountered various problems, always willing to cooperate with us, to us as the real God.
    5 Stars By Ophelia from Luxembourg - 2017.09.30 16:36

    Send your message to us:

    Write your message here and send it to us

    Send your message to us:

    Write your message here and send it to us